Abstract

The introduction of the ABSORB bioresorbable vascular scaffold (BVS) provides a new tool for stenting in interventional cardiology. Initially, relatively simple coronary artery lesions were treated with this novel device; nowadays, we are gaining clinical experience when treating a wide variety of lesions with the ABSORB BVS, including bifurcation lesions. Unfortunately, data are limited in terms of the use of the ABSORB BVS in coronary bifurcation lesions, so little is known about the safety and feasibility of these procedures. Bench testing and case reports showed that single provisional scaffold placement is feasible with fenestration of the scaffold towards the side branch and sequential non-compliant balloon inflation in the side and main branches. However, no prospective randomised clinical data with optical coherence tomography (OCT) imaging for different bifurcation stenting techniques are available. Based on the available data and our own experience we would recommend the use of the provisional single scaffold technique and only to fenestrate the scaffold if a severely pinched ostium combined with impaired flow seen on angiogram.

Disclosure:Joanna J Wykrzykowska receives consultancy fees from Abbott Vascular. The AMC Heartcenter received a restricted research grant from Abbott Vascular. The remaining authors have no conflicts of interest to declare.

Received:

Citation: Interventional Cardiology Review, 2014;9(2):84–8

Correspondence Details:Joanna J Wykrzykowska, Academic Medical Center – University of Amsterdam, Meibergdreef 9, 1105 AZ Amsterdam, The Netherlands. E: j.j.wykzykowska@amc.uva.nl

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

In the past three decades, significant progress has been made in the treatment of coronary artery disease. From the introduction of balloon angioplasty by Andreas Grüntzig in 19771 to metallic drugeluting stents with thin stent struts coated with cytotoxic or cytostatic drugs,2,3 with biocompatible/biodegradable polymers,4 with or without endothelial progenitor cell-capturing technology.5 However, current standard treatment with metallic stents has its shortcomings, such as late in-stent restenosis, late and very late stent thrombosis, impaired vasomotion in the stented segment6–8 and hindrance of repeat revascularisations. To potentially overcome the shortcomings of metallic stents, fully bioabsorbable stents (i.e. scaffolds) were developed.9 After preclinical testing and clinical evaluation in relatively simple coronary artery lesions,10–12 the ABSORB everolimus-eluting bioresorbable vascular scaffold (BVS) received CE-mark approval on 14 December 2010 and is since then increasingly being used in clinical practice across Europe and the rest of the world. This adoption in clinical practice led to a broad extension of, officially offlabel, indications in which the ABSORB BVS is being used, such as ST-segment elevation myocardial infarction (STEMI),13 chronic totally occluded (CTO) arteries and bifurcation lesions.

The use of bioresorbable technology in coronary bifurcation lesions may have potential benefits compared with metallic stents. For example, bioresorbable scaffolds could prevent permanent obstruction of a side branch (SB) after full absorption of the struts in front of this SB, leading to increased blood flow in the SB. Another potential benefit of the ABSORB BVS in bifurcation lesions might be a lower risk of late stent thrombosis, due to the absorption of non-apposed SB (NASB) struts at long term. These NASB struts are known to be often uncovered and thus a potential nidus for stent thrombosis.14 However, in all preclinical and clinical trials on the BVS, patients with SB ≥2 mm were excluded from enrolment, so limited data are available about the use of the BVS in bifurcation lesions. Therefore, the ‘Instructions for use’ of the ABSORB BVS does not indicate the use of the device in lesions involving a SB >2.5 mm. This led to some practical concerns about the use of the ABSORB BVS in coronary bifurcation lesions. For example, does the ABSORB BVS, with a strut thickness of 150 μm, allow for the same bifurcation techniques being used in metallic stent? Is the BVS, due to the lactic acid material, easier to fracture? This article will focus on data provided from in vitro and in vivo assessment of the ABSORB BVS in coronary bifurcation lesions and will provide practical recommendations based on these data and our own experience.

in vitro evaluation
Džavík and Colombo recently reported several bifurcation bench tests in which the ABSORB BVS is evaluated in a synthetic arterial model.15 Bifurcation stenting techniques tested included provisional stenting with final kissing balloon inflation (FKBI), modified T-stenting with a FKBI, double or two-step crush technique, mini-crush technique and culotte technique. The investigators used non-compliant (NC) balloons only, to prevent overexpansion of the scaffold due to increased balloon diameters. All procedures were assessed visually and by scanning electron microscopy (single scaffold techniques) and by microcomputed tomography (two scaffold techniques).

vitro Bench Test of the Double-crush Procedure
Download original

<>The investigators did not see any malapposition or strut fractures after FKBI in provisional stenting of a 3.0x18 mm BVS. FKBI in the provisional stenting technique was performed with a 2.5x20 mm balloon in the SB and 3.0x20 mm balloon in the main branch, both inflated at 8 atmospheres. Strut fractures were observed when FKBI was performed after T-stenting technique using a 3.0x18 mm BVS in the main branch (MB) and a 2.5x18 mm BVS in the SB and 3.0x20 mm (MB) and 2.5x20 mm (SB) balloons inflated at 10 atmospheres. The double- and mini-crush techniques both resulted in mild protrusions of BVS struts in the main branch. Even so, the double-crush technique showed a small area of malapposition between the carina and the two overlying scaffolds (see Figure 1). Likewise, a small area of malapposition was seen if the culotte technique was used, this technique also led to a thick circumferential two-layer scaffold wall in the proximal segment of the MB as well as a bulky BVS neocarina.

Based on their observations, the authors recommended to use provisional stenting in the majority of cases, with sequential noncompliant balloon inflation in the SB and the MB and reserving FKBI only if absolutely necessary. Restricted use of planned two-stent techniques, like crush or culotte, was recommended as more bench testing is needed to evaluate the feasibility of these techniques. These recommendations are supported by other bench test studies, performed by White et al. presented at Transcatheter Cardiovascular Therapeutics (TCT) 2013.sup>16

in vivo Evaluation
Systematic evaluation of side branches ≤2 mm jailed by the ABSORB BVS was performed by Okamura et al.17 Furthermore, few case reports described the treatment of true bifurcations lesions with involvement of SB ≥2 mm.18–21 To allow full assessment of the ABSORB BVS SB struts, optical coherence tomography (OCT) was used in all these reports. OCT allows visualisation of the vessel wall and the ABSORB BVS struts at high resolution22 and with a high degree of reproducibility.23

Okamura et al. assessed jailed side branch struts at baseline with three-dimensional (3D) OCT reconstruction, allowing the SB orifices to be evaluated visually. In total, 40 SB ostia of 25 patients enrolled in the ABSORB Cohort B study were analysed. In total, 24 orifices were jailed and showed different degrees and patterns of compartmentalisation (Type V, T and H) (see Figure 2).

Three-dimensional Assessment of Non-jailed and Jailed Side Branch
Download original
Optical Coherence Tomography Evaluation of a Left Main Bifurcation Lesion
Download original

Further insight in the use of the ABSORB BVS in coronary artery bifurcation lesions emerges from case reports combined with OCT images. Okamura et al. described a patient with an obstructed SB of ≤2 mm. Serial OCT imaging was performed at 2-year follow up showing an open SB ostium with intimal bridges creating a neocarina, persisting after absorption of the scaffold.24 Van Geuns et al. demonstrated the feasibility of crossing the ABSORB BVS struts into the SB with a coronary guidewire after deployment of a 3.0x18 mm ABSORB BVS in a left anterior descending (LAD) artery bifurcation lesion (Medina classification 1,1,0).21 Gogas et al. were first to describe side branch balloon dilatation of a LAD/second diagonal bifurcation lesion (Medina 1,1,1) treated with a single ABSORB BVS in the main branch. After deployment of a 3.0x18 mm ABSORB BVS in the midLAD, there was impaired flow and pinching of the ostium of the second diagonal, after which the scaffold was fenestrated towards the SB and dilated with a 1.5x12 mm Trek-compliant balloon. 2D OCT images and offline 3D-OCT reconstructions showed an open ostium of the SB and well-apposed struts in the SB segment.20 Finally, Grundeken et al. showed a successful deployment of a 3.5x18 mm ABSORB BVS in the left main coronary artery (LMCA) in the direction of the LAD, crossing the left circumflexus artery (LCx). However, after placement of the scaffold plaque shift towards the ostium of the LCx was observed on angiography. Therefore, the BVS was fenestrated towards the LCx and the LCx ostium was dilated with 2.0x20mm and 2.5x15mm Sapphire II compliant balloons at 8 atmospheres. Offline 3D-OCT reconstructions of the scaffold before fenestration showed multiple struts jailing the ostium of the LCx. After dilatation of the LCx ostium there is an opening of the stent cell seen towards the LCx without scaffold disruption (see Figure 3). These case reports illustrated that implantation of the ABSORB BVS and scaffold fenestration towards the SB with subsequent SB post-dilatation is feasible. Furthermore, these cases described the role of OCT to guide optimal scaffold placement.

Practical Recommendations
Although more clinical studies are needed to verify the optimal strategy in bifurcation lesions using ABSORB BVS, we would like to make the following recommendations based on previously described data and our own experience. In accordance with the recommendation of the European Bifurcation Club, based on randomised data on one- versus two-stent techniques using metallic stents, we would recommend the provisional single stenting technique in the majority of the lesions (see Figure 4), as previously suggested by Džavík and Colombo.15

Adequate lesion preparation and optimal sizing of the ABSORB BVS is essential prior to scaffold placement. Stent sizing with metallic stents is recommended to be based on the distal diameter with subsequent postdilatation of the proximal main branch with a larger balloon to ensure stent apposition in the proximal main branch (proximal optimisation technique [POT]).25 This technique is considered to prevent carinal shift and subsequent ostial pinching of the SB. However, postdilation of the BVS with balloons >0.5 mm larger than the BVS size is discouraged due to the increased risk of strut fractures. Therefore, we recommend to choose the BVS size based on the proximal diameter. Intravascular imaging, such as intravascular ultrasound (IVUS) or OCT, or online quantitative coronary angiography (QCA) can be used for optimal sizing of the ABSORB BVS.

Provisional Single-scaffold Placement in a Coronary Bifurcation Lesion
Download original
Provisional Single-scaffold Placement in a Coronary Bifurcation Lesion
Download original

After deployment of the scaffold, we recommend to guide further need for fenestration and post-dilatation of the scaffold on the angiographic result and OCT images. If a lesion after deployment of the scaffold has a severely pinched ostium on angiography combined with impaired flow, Thrombolysis In Myocardial Infarction (TIMI) flow 1 or 2, we recommend the fenestration of the scaffold towards the SB. Online 3D-OCT reconstructions, which are available at the newest OCT consoles, could guide the optimal guidewire position during recrossing towards the SB. Recrossing through the most distal compartment is recommended, as distal recrossing leads to a reduction of incomplete stent apposition (ISA) at baseline in metallic stents.26 In our experience, fenestration of the scaffold towards the SB is feasible with a 2.0 mm or a 2.5 mm balloon in the SB ostium to a maximum of 6 to 8 atmospheres. Final OCT pullback after fenestration could be performed to detect whether the scaffold is well apposed in the whole stented area and to control for any struts out of the expected range. The latter is a sign of strut fracture, which could be treated with additional stenting. Due to lack of clinical data, we recommend repeat angiography to be performed combined with OCT at 9 months after scaffold placement, even if the patient is without any complaints. This allows assessment of the TIMI flow in the SB and MB, intimal tissue bridging in front of the SB and intimal formation in the distal MB. Treatment of ABSORB BVS re-stenosis has not been investigated in clinical trials; however, Nakatani et al. showed a case series of six patients with in-stent restenosis who have been treated with additional stenting using Xience drug-eluting stents.27{C}

Discussion
The ABSORB BVS is currently being used in daily clinical practice around the world. Unfortunately, very limited data are available about the use of the BVS in coronary bifurcation lesions. Several case reports have shown that treatment of bifurcation lesions and fenestration of the scaffold towards the side branch is feasible, which is supported by in vitro testing. However, all recommendations made in the treatment of coronary bifurcation lesions are not supported by any long-term clinical or in vivo data and are solely based on either expert opinion or limited in vitro data.

Džavík and Colombo are the first to publish in vitro benchmark testing. The tests performed are extensive; however, only one or two experiments were performed in each bifurcation stenting procedure. For example, a double-crush procedure could be successful once or twice but what if the experiment is repeated multiple times, will these results still hold? Although silicone models provide us with valuable insights into BVS geometry, in vitro models almost always have their limitations, which makes it difficult to deduce these results in terms of in vivo use. Silicone models are in general more rigid compared with coronary arteries, which makes it harder to overstretch and fracture the ABSORB BVS. Furthermore, the radial force of these models differ from coronary arteries, which makes assessment of non-apposed struts less reliable.

Okamura et al. provided a detailed description of how the ABSORB BVS obstructs SB orifices at baseline and the patterns of compartmentalisation and the number of compartments obstructing the SB. These observations are intellectually interesting; however, it is uncertain what implications it has for everyday clinical practice. Furthermore, the clinical outcome in relation with the pattern of compartmentalisation and number of compartments is still unknown and needs further investigation.

Moreover, the effect of different bifurcation stenting procedures at follow up is unknown. Different stenting procedures could alter the endothelial shear stress patterns and coronary blood flow in different ways, hypothetically leading to different neointimal formation patterns at 1-year follow up.28 The effect of all these factors could alter the neo-intimal bridge formation in front of the side branch in two ways. At follow-up this could potentially lead to increased neo-intimal formation with obstruction of the SB, although also non-covered NASB struts can persist.

Due to the lack of data, we believe that there is an urgent need for prospective randomised clinical data with OCT imaging for different bifurcation stenting techniques using the ABSORB BVS. In the absence of these data we recommend the use of the single scaffold provisional stenting strategy in coronary artery bifurcation lesions as is recommended by the European bifurcation club for metallic stents.

Conclusion
Based on case reports and in vitro testing ABSORB BVS placement in coronary bifurcation lesions seems to be feasible. The potential theoretical advantage of BVS of better access to side branches due to the disappearance over time of obstructing struts needs to be confirmed. Complex multiple BVS-bifurcation constructions, such as Culotte or Crush, are unattractive due to mechanical properties and strut size of currently available devices. Further research should be performed to detect the optimal bifurcation stenting procedure for this novel devices.

References

  1. Gruntzig A, Transluminal dilatation of coronary-artery stenosis, Lancet, 1978;1:263.
  2. Stone GW, Rizvi A, Newman W, et al., Everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease, N Engl J Med, 2010;362:1663–74.
  3. Stefanini GG, Kalesan B, Serruys PW,et al., Long-term clinical outcomes of biodegradable polymer biolimus-eluting stents versus durable polymer sirolimus-eluting stents in patients with coronary artery disease (LEADERS): 4 year follow-up of a randomised non-inferiority trial, Lancet, 2011;378:1940–8.
  4. Separham A, Sohrabi B, Aslanabadi N, Ghaffari S, The twelvemonth outcome of biolimus eluting stent with biodegradable polymer compared with an everolimus eluting stent with durable polymer, J Cardiovasc Thorac Res, 2011;3:113–6.
  5. Woudstra P, de Winter RJ, Beijk MA, Next-generation DES: the COMBO dual therapy stent with Genous endothelial progenitor capturing technology and an abluminal sirolimus matrix, Expert Rev Med Devices, 2014;11:121–35.
  6. Nakazawa G, Otsuka F, Nakano M, et al., The pathology of neoatherosclerosis in human coronary implants bare-metal and drug-eluting stents, J Am Coll Cardiol, 2011;57:1314–22.
  7. Joner M, Finn AV, Farb A, et al., Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk, J Am Coll Cardiol, 2006;48:193–202.
  8. Maier W, Windecker S, Kung A, et al., Exercise-induced coronary artery vasodilation is not impaired by stent placement, Circulation, 2002;105:2373–7.
  9. Oberhauser JP, Hossainy S, Rapoza RJ, Design principles and performance of bioresorbable polymeric vascular scaffolds. EuroIntervention, 2009;5 Suppl. F:F15–F22.
  10. Serruys PW, Ormiston JA, Onuma Y, et al., A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods, Lancet, 2009;373:897–910.
  11. Serruys PW, Onuma Y, Ormiston JA, et al., Evaluation of the second generation of a bioresorbable everolimus drugeluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes, Circulation, 2010;122:2301–12.
  12. Serruys PW, Onuma Y, Dudek D, et al.,Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes, J Am Coll Cardiol, 2011;58:1578–88.
  13. Kajiya T, Liang M, Sharma RK, et al., Everolimus-eluting bioresorbable vascular scaffold (BVS) implantation in patients with ST-segment elevation myocardial infarction (STEMI), EuroIntervention, 2013;9:501-504.
  14. Gutierrez-Chico JL, Regar E, Nuesch E, et al., Delayed coverage in malapposed and side-branch struts with respect to well-apposed struts in drug-eluting stents: in vivo assessment with optical coherence tomography, Circulation, 2011;124:612–23.
  15. Džavík V, Colombo A, The absorb bioresorbable vascular scaffold in coronary bifurcations: insights from bench testing. JACC Cardiovasc Interv, 2014;7:81–8.
  16. White JM, Bifurcation strategies with the Absorb BVS Resorbable Scaffold. In: TCT, San Francisco, CA, USA, 2013.
  17. Okamura T, Onuma Y, Garcia-Garcia HM, et al., 3-dimensional optical coherence tomography assessment of jailed side branches by bioresorbable vascular scaffolds: a proposal for classification. JACC Cardiovasc Interv, 2010;3:836–44.
  18. Grundeken MJ, Kraak RP, de Bruin DM, Wykrzykowska JJ, Three-dimensional optical coherence tomography evaluation of a left main bifurcation lesion treated with ABSORB(R) bioresorbable vascular scaffold including fenestration and dilatation of the side branch, Int J Cardiol, 2013;168e107–e108.
  19. Dzavik V, Muramatsu T, Crooks N, et al., Complex bifurcation percutaneous coronary intervention with the Absorb bioresorbable vascular scaffold. EuroIntervention, 2013;9:888.
  20. Gogas BD, van Geuns RJ, Farooq V, et al., Three-dimensional reconstruction of the post-dilated ABSORB everolimus-eluting bioresorbable vascular scaffold in a true bifurcation lesion for flow restoration. JACC Cardiovasc Interv, 2011;4:1149–50.
  21. van Geuns RJ, Gogas BD, Farooq V, et al., 3-dimensional reconstruction of a bifurcation lesion with double wire after implantation of a second generation everolimuseluting bioresorbable vascular scaffold, Int J Cardiol, 2011;153:e43–e45.
  22. Gutierrez-Chico JL, Serruys PW, Girasis C, et al., Quantitative multi-modality imaging analysis of a fully bioresorbable stent: a head-to-head comparison between QCA, IVUS and OCT, Int J Cardiovasc Imaging, 2012;28:467–78.
  23. Gomez-Lara J, Brugaletta S, Diletti R, et al., Agreement and reproducibility of gray-scale intravascular ultrasound and optical coherence tomography for the analysis of the bioresorbable vascular scaffold, Catheter Cardiovasc Interv, 2012;79:890–902.
  24. Okamura T, Serruys PW, Regar E, Cardiovascular flashlight. The fate of bioresorbable struts located at a side branch ostium: serial three-dimensional optical coherence tomography assessment, Eur Heart J, 2010;31:2179.
  25. Stankovic G, Lefevre T, Chieffo A, et al., Consensus from the 7th European Bifurcation Club meeting, EuroIntervention, 2013;9:36–45.
  26. Okamura T, Onuma Y, Yamada J, et al., 3D optical coherence tomography: new insights into the process of optimal rewiring of side branches during bifurcational stenting, EuroIntervention, 2014;pii: 20130514–02.
  27. Nakatani S, Onuma Y, Ishibashi Y, et al., Early (before 6 months), late (6-12 months) and very late (after 12 months) angiographic scaffold restenosis in the ABSORB Cohort B trial, EuroIntervention, 2014;pii: 20130829-09..
  28. Bourantas CV, Papafaklis MI, Kotsia A, et al., Effect of the endothelial shear stress patterns on neointimal proliferation following drug-eluting bioresorbable vascular scaffold implantation: an optical coherence tomography study, JACC Cardiovasc Interv, 2014;7:315–24.